5.1. TECHNICAL SPECIFICATIONS
1. Capacities: 150 lb/68 kg factory set
2. Rounding: Nearest division per NIST H-44
3. Weight Display: .27 inch LCD Display
4. Power On Lockout: Scale will display "------", when power is turned on and
weight is present. Press ZERO to establish zero reference.
5. Display update rate: 0.05 seconds.
6. Overcapacity Warning:
Displays "HiCAP" for overcapacity
7. Motion Detection: Satisfies H-44 requirements
8. Power Failure Protection: Zero reference, programming, and calibrations are
retained if the instrument loses power
9. Load Cell Excitation: 3.3VDC
10. USB Cable Length: Type A/B, 6 feet in length
11. Indicators:
lb, kg
12. Dimensions: Platform, Including Feet 14" x 14" x 2.5"
13. Auto Zero Tracking: Compensates for gradual buildup of material on
platform, factory set at 0.5 divisions
14. Power Requirements: 5VDC, USB or AC adapter
15. Approvals: NTEP CC# 98-198
MC# AM-5298
RoHS Compliant