Section 5: Commissioning
05/11 44 51209 Revision 2
5.4.17 Test Value Determination/Display
Press [Test] to activate test value calculation. The maximum capacity (Max) is displayed with the
designation without a weight unit.
WP A/Calibration
Test
Depending on the setting under -[Calibration]-[Param]-[Test mode], either is shown when you
press
to view the test data.
- the maximum capacity with [Absolute], or
- the deviation from the test value with [Relative]
5.4.18 Finishing/Saving the Calibration
Finish the calibration with .
You are prompted to confirm whether calibration should be closed
without determining the test value.
Calibration not complete
Exit calibration?
If not data all was determined when calibrating with [New] (e.g. dead
load not set/entered), this message is shown.
Press [Yes] to confirm and then press again; another prompt is
displayed:
Exit calibration ?
Press [Save] to save changes in calibration data.
If you press [Undo], changes are not saved and the display returns to the
selection menu for the weighing points.
Saving calibration Confirm, or
Exit calibration close the menu.
After finishing the calibration, set the CAL switch to the closed position; see also Chapter 5.1.1.